FDA Adverse Event
Malfunction
Summary report: N
I-STAT TROPONIN (CTNI) CARTRIDGE
MDR report key: 2192508
·
Received June 30, 2011
Report
- Report Number
- 2245578-2011-00109
- Event Type
- Malfunction
- Date Received
- June 30, 2011
- Date of Event
- April 15, 2011
- Report Date
- June 30, 2011
- Manufacturer
- ABBOTT POINT OF CARE CANADA
- Product Code
- MMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). INVESTIGATION COMPLETED ON (B)(4) 2011, THAT CONCLUDED THE FOLLOWING: RESULTS FROM FINISHED GOODS AND RETURNED CARTRIDGE TESTING INDICATE THAT CTNI LOT (B)(4) CARTRIDGES HAD A SECTION OF 250 CARTRIDGES OUT OF A TOTAL OF 8800 CARTRIDGES WERE FOUND TO BE DEFICIENT WITH RESPECT TO THE RATE QCC 123 RESULTS. THERE WERE NO OTHER DEFICIENCIES OR QUALITY ISSUES IDENTIFIED RELATED TO THE CUSTOMERS COMPLAINT.
Description of Event or Problem · 1
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT WERE YIELDING QUALITY CHECK CODES 123, 133, AND 143. BASED ON THE INFO AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION WAS COMPLETED ON (B)(4) 2011.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE CANADA | NA | T11010 |