FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 2192508 · Received June 30, 2011

Report

Report Number
2245578-2011-00109
Event Type
Malfunction
Date Received
June 30, 2011
Date of Event
April 15, 2011
Report Date
June 30, 2011
Manufacturer
ABBOTT POINT OF CARE CANADA
Product Code
MMI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED ON (B)(4) 2011, THAT CONCLUDED THE FOLLOWING: RESULTS FROM FINISHED GOODS AND RETURNED CARTRIDGE TESTING INDICATE THAT CTNI LOT (B)(4) CARTRIDGES HAD A SECTION OF 250 CARTRIDGES OUT OF A TOTAL OF 8800 CARTRIDGES WERE FOUND TO BE DEFICIENT WITH RESPECT TO THE RATE QCC 123 RESULTS. THERE WERE NO OTHER DEFICIENCIES OR QUALITY ISSUES IDENTIFIED RELATED TO THE CUSTOMERS COMPLAINT.

Description of Event or Problem · 1

ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT WERE YIELDING QUALITY CHECK CODES 123, 133, AND 143. BASED ON THE INFO AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION WAS COMPLETED ON (B)(4) 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE CANADA NA T11010

Patients

Seq Age Sex Outcome Treatment
1