ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05636
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- July 10, 2011
- Report Date
- July 10, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
CORRECTION NUMBER: 2531779-03/24/2010-003-R. THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS NOT DUPLICATED DURING TESTING. UNRELATED TO THE COMPLAINT, THE BATTERY CAP WAS FOUND TO BE STRIPPED AND THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED. THE USER GUIDE INSTRUCTS THE PATIENT TO REPLACE THE BATTERY CAP AT LEAST ONCE PER YEAR AND THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND / OR THE WATERPROOF FEATURE OF THE PUMP.
THE PATIENT REPORTED THAT THE PUMP PUSHED ALL OF THE INSULIN OUT OF THE CARTRIDGE DURING THE LOAD STEP OF A ROUTINE CARTRIDGE CHANGE. SHE STATED THAT SHE ATTEMPTED TO LOAD ANOTHER CARTRIDGE, AND THE SAME ISSUE OCCURRED. THERE WAS NO REPORTED PATIENT IMPACT AS A RESULT OF THE INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |