MDR report key: 2192484 · Received July 14, 2011

Report

Report Number
1929691-2011-00007
Event Type
Injury
Date Received
July 14, 2011
Date of Event
June 6, 2011
Report Date
June 30, 2011
Manufacturer
STANLEY SECURITY SOLUTIONS, INC., SENIOR TECHNOLOGIES DIVISION
Product Code
ILQ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6) INDICATED THAT A RESIDENT HAD FALLEN IN HIS/HER BATHROOM (C3) ON (B)(6) 2011 AND THAT STAFF AT (B)(6) CLAIMED TO NOT HAVE RECEIVED PAGES ASSOCIATED WITH THE INCIDENT. (B)(4) TECHNICIANS WERE AT (B)(6) ON (B)(4) 2011 AND COULD NOT DURING ANY OF THE VISITS REPLICATE THE ISSUES AS DESCRIBED. AFTER EXTENSIVE TESTING AND REVIEW OF THE CUSTOMERS RECORDS, THE FOLLOWING WAS DETERMINED: THE SYSTEM MONITOR (COMPUTER) CLEARLY RECORDED THE INCIDENT CALL AND THE LATER "CANCEL" OF THE CALL. THE RESIDENT IN QUESTION REGISTERED 11 ALARM CALLS TO THE ARIAL SYSTEM ON THE DAY OF THE INCIDENT ((B)(6) 2011). THE AVERAGE STAFF RESPONSE TIME TO THAT RESIDENT THAT DAY WAS 8.54 MINUTES. THE LONGEST CALL INTERVAL ASIDE FROM THE INCIDENT CALL WAS 12.5 MINUTES; THE SHORTEST WAS THE FACILITY'S TEST CALL IMMEDIATELY FOLLOWING THE INCIDENT. DISREGARDING THE SHORTEST AND LONGEST CALLS, THE STAFF'S AVERAGE RESPONSE TIME TO THAT RESIDENT THAT DAY WAS 6 MINUTES. (THERE WERE ALSO THREE SAME DAY CALLS AFTER THE INCIDENT FROM THE RESIDENT'S ROOM.) DURING THE 39 MINUTE INTERVAL OF THE INCIDENT ALARM, THERE WERE 6 OTHER CALLS TO THE ARIAL SYSTEM, WHICH RESULTED IN APPROXIMATELY 30 SUCCESSFUL NOTIFICATIONS TO STAFF PAGERS (AS EVIDENCED BY THE CALL "CANCEL" TIMES ON THE SYSTEM). CORRECT OPERATION OF THE SYSTEM WAS VERIFIED BY RESPONSE TO OTHER CALLS DURING THE SAME TIME PERIOD. (B)(4).

Description of Event or Problem · 1

ON (B)(6) 2011, (B)(6) CONTACTED (B)(4) TECHNICAL SERVICE DEPARTMENT AND CLAIMED THAT THE ARIAL SYSTEM STOPPED DISPLAYING ROOM CALLS WHILE TESTING THEIR FIRE SYSTEM. SUBSEQUENT TO THIS CALL, A (B)(4) TECHNICIAN RESPONDED TO THE REQUEST FOR REPAIR ON (B)(6) 2011. DURING OUR TECHNICIAN'S VISIT ON (B)(4) 2011, IT WAS MENTIONED BY (B)(6) THAT HER RESIDENT HAD FALLEN IN HIS/HER BATHROOM (C3) ON (B)(6) 2011, AND THAT STAFF AT (B)(6) CLAIMED TO NOT HAVE RECEIVED PAGES ASSOCIATED WITH THE INCIDENT. ACCORDING TO THE ARIAL REPORTS AND THE INFORMATION STAFF PROVIDED, THE RESIDENT'S ARIAL WALL TRANSMITTER HAD BEEN ACTIVATED AND HAD REMAINED ACTIVE FOR 39 MINUTES STARTING AT 1:45 PM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARIAL WALL TRANSMITTER, PAGER ILQ STANLEY SECURITY SOLUTIONS, INC., SENIOR TECHNOLOGIES DIVISION 54037 NA