FLEXCATH STEERABLE SHEATH
MDR report key: 2192441
·
Received August 5, 2011
Report
- Report Number
- 3002648230-2011-00115
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Date of Event
- July 6, 2011
- Report Date
- July 6, 2011
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- DRA
- PMA / PMN Number
- K081049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN (B)(4) 2011, TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.
Description of Event or Problem · 1
DURING A CRYOABLATION PROCEDURE, AFTER THE REMOVAL OF A MAPPING CATHETER FROM THE FLEXCATH STEERABLE SHEATH AND THE INTRODUCTION OF THE ARCTIC FRONT CATHETER, AIR WAS ASPIRATED INTO THE SHEATH. THE FLEXCATH STEERABLE SHEATH WAS REPLACED AND NO ADVERSE EVENT OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXCATH STEERABLE SHEATH | CATHETER, STEERABLE | DRA | MEDTRONIC CRYOCATH LP | 3FC12 | 38075 |