FDA Adverse Event Malfunction Summary report: N

FLEXCATH STEERABLE SHEATH

MDR report key: 2192441 · Received August 5, 2011

Report

Report Number
3002648230-2011-00115
Event Type
Malfunction
Date Received
August 5, 2011
Date of Event
July 6, 2011
Report Date
July 6, 2011
Manufacturer
MEDTRONIC CRYOCATH LP
Product Code
DRA
PMA / PMN Number
K081049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN (B)(4) 2011, TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.

Description of Event or Problem · 1

DURING A CRYOABLATION PROCEDURE, AFTER THE REMOVAL OF A MAPPING CATHETER FROM THE FLEXCATH STEERABLE SHEATH AND THE INTRODUCTION OF THE ARCTIC FRONT CATHETER, AIR WAS ASPIRATED INTO THE SHEATH. THE FLEXCATH STEERABLE SHEATH WAS REPLACED AND NO ADVERSE EVENT OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXCATH STEERABLE SHEATH CATHETER, STEERABLE DRA MEDTRONIC CRYOCATH LP 3FC12 38075

Patients

Seq Age Sex Outcome Treatment
1