OT ULTRALINK METER
Report
- Report Number
- 2939301-2011-06740
- Event Type
- Malfunction
- Date Received
- August 5, 2011
- Report Date
- July 16, 2011
- Manufacturer
- LIFESCAN INC.
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- OTHER
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K)# IS K073231.
FOLLOW-UP # 1 (09/08/2011)-DEVICE EVALUATION: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6), 2011 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS HAVE PASSED ALL TESTING WITH NO FAULTS FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THERE IS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE INACCURATE ERRATIC WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT ULTRALINK METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN INC. | 3145135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |