FDA Adverse Event Malfunction Summary report: N

ADULT DUAL-HEATED BREATHING CIRCUIT

MDR report key: 2192367 · Received August 4, 2011

Report

Report Number
9611451-2011-00471
Event Type
Malfunction
Date Received
August 4, 2011
Date of Event
June 16, 2011
Report Date
July 5, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
REFER TO H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT204 ADULT BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS (B)(4). METHOD: THE PINS ON THE EXPIRATORY AND INSPIRATORY LIMB OF THE COMPLAINT BREATHING CIRCUIT WERE TESTED TO SEE IF THEY COULD BE CONNECTED TO A HEATER WIRE ADAPTOR. RESULTS: THE HEATER WIRE PINS OF THE INSPIRATORY LIMB WERE WITHIN SPECIFICATION FOR THIS PRODUCT AS FULL INSERTION OF THE HEATER WIRE ADAPTOR WAS ACHIEVED. THE HEATER WIRE PINS OF THE EXPIRATORY LIMB WERE OUT OF SPECIFICATION AS A HEATER WIRE PIN WAS SPLIT, THEREFORE FULL INSERTION OF THE HEATER WIRE ADAPTOR WAS NOT POSSIBLE. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. IT IS POSSIBLE FOR THE USER TO DAMAGE THE HEATER WIRE PINS DURING USE IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. FOR DAMAGED PINS REPORTED TO US BY HEALTHCARE FACILITIES, IT IS IMPOSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE DAMAGED DURING PRODUCTION OR BY THE END USER. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE HAS RECENTLY BEEN RETURNED TO FISHER & PAYKEL HEALTHCARE AND INVESTIGATION IS ANTICIPATED. WE WILL SEND A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE EXPIRATORY LIMB HEATER WIRE OF AN (B)(4) ADULT DUAL-HEATED BREATHING CIRCUIT COULD NOT BE CONNECTED. THIS WAS FOUND PRIOR TO USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE EXPIRATORY LIMB HEATER WIRE OF AN RT204 ADULT DUAL-HEATED BREATHING CIRCUIT COULD NOT BE CONNECTED. THIS WAS FOUND PRIOR TO USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL-HEATED BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LTD RT204 100129

Patients

Seq Age Sex Outcome Treatment
1