FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2192310
·
Received August 4, 2011
Report
- Report Number
- 2050012-2011-03900
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Date of Event
- July 7, 2011
- Report Date
- July 7, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BEC FIELD SERVICE ENGINEER (FSE) REPLACED A DEFECTIVE VALVE ON THE FLUID DISTRIBUTION MANIFOLD THAT CAUSED THE LEAK. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) TO REPORT OBSERVING SLOW DRIPPING LEAK FROM REAGENT PROBE A. THE CUSTOMER WORE PPE TO CLEAN THE SPILL WITH GAUZE. NO RESULTS HAVE BEEN AFFECTED BY THIS EVENT. NO INJURY OR EXPOSURE TO FLUID WAS REPORTED
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE. | JJE | BECKMAN COULTER INC. | DXC 600I | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |