FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2192310 · Received August 4, 2011

Report

Report Number
2050012-2011-03900
Event Type
Malfunction
Date Received
August 4, 2011
Date of Event
July 7, 2011
Report Date
July 7, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) REPLACED A DEFECTIVE VALVE ON THE FLUID DISTRIBUTION MANIFOLD THAT CAUSED THE LEAK. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) TO REPORT OBSERVING SLOW DRIPPING LEAK FROM REAGENT PROBE A. THE CUSTOMER WORE PPE TO CLEAN THE SPILL WITH GAUZE. NO RESULTS HAVE BEEN AFFECTED BY THIS EVENT. NO INJURY OR EXPOSURE TO FLUID WAS REPORTED