FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2192158
·
Received August 4, 2011
Report
- Report Number
- 2050012-2011-03892
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Date of Event
- July 5, 2011
- Report Date
- July 6, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BEC FIELD SERVICE ENGINEER REPLACED LINES 142, 145, &127 TO CORRECT THE WASH CONCENTRATE LEAK. BEC INTERNAL NOTIFICATION NUMBER IS (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT OBSERVING YELLOW DRIED CRYSTALS ON THE BOTTOM OF THE HYDRO-PNEUMATIC DRAWER AND FLUID LEAK. THE CUSTOMER WAS UNABLE TO IDENTIFY THE SOURCE OF THE LEAK. NO INJURY OR EXPOSURE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | DXC 600 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |