FDA Adverse Event Malfunction Summary report: N

SUBQ

MDR report key: 2192121 · Received August 4, 2011

Report

Report Number
2182208-2011-01081
Event Type
Malfunction
Date Received
August 4, 2011
Date of Event
July 21, 2011
Manufacturer
MEDTRONIC, INC.
Product Code
LWS
PMA / PMN Number
P920015/S021
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE DEFIBRILLATION COIL WAS DISTORTED. IT WAS NOTED THAT THE LEAD APPEARED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE COIL WAS BENT AND THE LEAD WAS POSSIBLY DAMAGED. THE LEAD WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUBQ IMPLANTABLE TACHY LEAD LWS MEDTRONIC, INC. 6996 ASKU

Patients

Seq Age Sex Outcome Treatment
1 7 YR Other