FDA Adverse Event Malfunction Summary report: N

RF3000 RADIOFREQUENCY GENERATOR

MDR report key: 2192069 · Received August 4, 2011

Report

Report Number
3005099803-2011-02633
Event Type
Malfunction
Date Received
August 4, 2011
Report Date
July 13, 2011
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
GEI
PMA / PMN Number
K000241
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PATIENT IDENTIFIER, AGE/DATE OF BIRTH, GENDER, AND WEIGHT ARE UNKNOWN. THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS; THEREFORE, THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. (B)(4).

Additional Manufacturer Narrative · 1

ENVIRONMENTAL STRESS TESTING, UNDER CONDITIONS OF HIGH AND LOW TEMPERATURE, HIGH AND LOW MARGIN (SUPPLY) VOLTAGE, AND SHORT-CIRCUIT SHUTDOWN, WAS PERFORMED. THE GENERATOR DID NOT EXHIBIT ANY MALFUNCTION WHICH COULD ACCOUNT FOR THE EVENT. FURTHERMORE, THE DEVICE PASSED ALL PERFORMANCE CRITERIA OF ITS FINAL TEST PROCEDURE.THE CONDITION OF THE RETURNED INCIDENT DEVICE SHOWED NO EVIDENCE OF EITHER THE ALLEGED ISSUE OR ANY DEFECT WHICH COULD HAVE CONTRIBUTED TO THE EVENT. THE COMPLAINT WAS NOT CONFIRMED.A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED.(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RF 3000 RADIOFREQUENCY GENERATOR WAS USED DURING A RFA (RADIOFREQUENCY ABLATION) PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A GENERATOR ERROR (P4) OCCURRED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RF 3000 RADIOFREQUENCY GENERATOR WAS USED DURING A RFA (RADIOFREQUENCY ABLATION) PROCEDURE.ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A GENERATOR ERROR (P4) OCCURRED.THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.