RF3000 RADIOFREQUENCY GENERATOR
Report
- Report Number
- 3005099803-2011-02633
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Report Date
- July 13, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MARLBOROUGH
- Product Code
- GEI
- PMA / PMN Number
- K000241
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
PATIENT IDENTIFIER, AGE/DATE OF BIRTH, GENDER, AND WEIGHT ARE UNKNOWN. THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS; THEREFORE, THE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. (B)(4).
ENVIRONMENTAL STRESS TESTING, UNDER CONDITIONS OF HIGH AND LOW TEMPERATURE, HIGH AND LOW MARGIN (SUPPLY) VOLTAGE, AND SHORT-CIRCUIT SHUTDOWN, WAS PERFORMED. THE GENERATOR DID NOT EXHIBIT ANY MALFUNCTION WHICH COULD ACCOUNT FOR THE EVENT. FURTHERMORE, THE DEVICE PASSED ALL PERFORMANCE CRITERIA OF ITS FINAL TEST PROCEDURE.THE CONDITION OF THE RETURNED INCIDENT DEVICE SHOWED NO EVIDENCE OF EITHER THE ALLEGED ISSUE OR ANY DEFECT WHICH COULD HAVE CONTRIBUTED TO THE EVENT. THE COMPLAINT WAS NOT CONFIRMED.A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED.(B)(4).
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RF 3000 RADIOFREQUENCY GENERATOR WAS USED DURING A RFA (RADIOFREQUENCY ABLATION) PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A GENERATOR ERROR (P4) OCCURRED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RF 3000 RADIOFREQUENCY GENERATOR WAS USED DURING A RFA (RADIOFREQUENCY ABLATION) PROCEDURE.ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A GENERATOR ERROR (P4) OCCURRED.THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RF3000 RADIOFREQUENCY GENERATOR | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | BOSTON SCIENTIFIC - MARLBOROUGH | M001262200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |