FDA Adverse Event Injury Summary report: N

MOMENTUM EL ICD DR

MDR report key: 21920552 · Received April 28, 2025

Report

Report Number
2124215-2025-26917
Event Type
Injury
Date Received
April 28, 2025
Date of Event
April 22, 2025
Report Date
April 16, 2026
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
LWS
UDI-DI
00802526587733
PMA / PMN Number
P960040/S385
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT WAS AMENDED TO UPDATE INTERNATIONAL MEDICAL DEVICE REGULATORS FORUM (IMDRF) CODES. INCOMPLETE CODING OF THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS ASSOCIATED WITH CAPA-8580, WHICH FOCUSED ON IDENTIFYING AND REMEDIATING INCOMPLETE CODING OF EVENTS. ADDED PATIENT CODE E210401. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED. THE REVIEW OF THE DHR IDENTIFIED THAT THERE WERE NO PROCESS RELATED NON-CONFORMANCES, SCRAP, OR REWORK PERFORMED DURING THE PRODUCTION THAT COULD EXPLAIN THE EVENT. THE REVIEWS ENSURE EACH DEVICE MEETS SPECIFICATION PRIOR TO RELEASE FOR USE. THERE IS NO INDICATION THE DEVICE MANUFACTURING PROCESS CONTRIBUTED TO THE REPORTED COMPLAINT. A RISK REVIEW WAS COMPLETED AND CONFIRMED THAT THE EVENT OF INAPPROPRIATE SHOCK WAS DEFINED IN THE RISK DOCUMENTATION. THIS EVENT TYPE HAS BEEN ACCOUNTED FOR DURING PRODUCT RISK ANALYSIS TO SUPPORT ACCEPTABLE RISK BENEFIT FOR THE PRODUCT. REVIEW OF LABELING DETERMINED THAT THE COMPLAINT SITUATION WAS LISTED IN THE MANUAL. THERE WAS NO INDICATION IN THE COMPLAINT THAT THE PRODUCT WAS NOT USED IN ACCORDANCE TO LABELING. THE MANUAL WAS UNLIKELY TO BE THE CAUSE OF THE REPORTED COMPLAINT; TRANSLATION, WORDING, OR GRAPHICS DOES NOT REQUIRE FURTHER REVIEW. THE DEVICE REMAINS IMPLANTED AND IN SERVICE; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. BASED ON A THOROUGH REVIEW OF THE REPORTED COMPLAINT, BOSTON SCIENTIFIC HAS ASSIGNED AN INVESTIGATION CONCLUSION CODE OF CAUSE NOT ESTABLISHED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE SHOCK THERAPY FOR ATRIAL FIBRILLATION (AF). THERAPY WAS EXHAUSTED. IT WAS NOTED THAT TWO SHOCKS TRIGGERED A CODE 1005, INDICATING AN OPEN CIRCUIT CONDITION. TECHNICAL SERVICES (TS) REVIEWED THE REPORTS AND PROVIDED TROUBLESHOOTING ACTIONS. THE DEVICE REMAINS IN USE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE SHOCK THERAPY FOR ATRIAL FIBRILLATION (AF). THERAPY WAS EXHAUSTED. IT WAS NOTED THAT TWO SHOCKS TRIGGERED A CODE 1005, INDICATING AN OPEN CIRCUIT CONDITION. TECHNICAL SERVICES (TS) REVIEWED THE REPORTS AND PROVIDED TROUBLESHOOTING ACTIONS. THE DEVICE REMAINS IN USE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341990 MOMENTUM EL ICD DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) LWS BOSTON SCIENTIFIC CORPORATION D121 240942 00802526587733

Patients

Seq Age Sex Outcome Treatment
1 90 YR Unknown Other