MOMENTUM EL ICD DR
Report
- Report Number
- 2124215-2025-26917
- Event Type
- Injury
- Date Received
- April 28, 2025
- Date of Event
- April 22, 2025
- Report Date
- April 16, 2026
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- LWS
- UDI-DI
- 00802526587733
- PMA / PMN Number
- P960040/S385
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS REPORT WAS AMENDED TO UPDATE INTERNATIONAL MEDICAL DEVICE REGULATORS FORUM (IMDRF) CODES. INCOMPLETE CODING OF THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS ASSOCIATED WITH CAPA-8580, WHICH FOCUSED ON IDENTIFYING AND REMEDIATING INCOMPLETE CODING OF EVENTS. ADDED PATIENT CODE E210401. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED. THE REVIEW OF THE DHR IDENTIFIED THAT THERE WERE NO PROCESS RELATED NON-CONFORMANCES, SCRAP, OR REWORK PERFORMED DURING THE PRODUCTION THAT COULD EXPLAIN THE EVENT. THE REVIEWS ENSURE EACH DEVICE MEETS SPECIFICATION PRIOR TO RELEASE FOR USE. THERE IS NO INDICATION THE DEVICE MANUFACTURING PROCESS CONTRIBUTED TO THE REPORTED COMPLAINT. A RISK REVIEW WAS COMPLETED AND CONFIRMED THAT THE EVENT OF INAPPROPRIATE SHOCK WAS DEFINED IN THE RISK DOCUMENTATION. THIS EVENT TYPE HAS BEEN ACCOUNTED FOR DURING PRODUCT RISK ANALYSIS TO SUPPORT ACCEPTABLE RISK BENEFIT FOR THE PRODUCT. REVIEW OF LABELING DETERMINED THAT THE COMPLAINT SITUATION WAS LISTED IN THE MANUAL. THERE WAS NO INDICATION IN THE COMPLAINT THAT THE PRODUCT WAS NOT USED IN ACCORDANCE TO LABELING. THE MANUAL WAS UNLIKELY TO BE THE CAUSE OF THE REPORTED COMPLAINT; TRANSLATION, WORDING, OR GRAPHICS DOES NOT REQUIRE FURTHER REVIEW. THE DEVICE REMAINS IMPLANTED AND IN SERVICE; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. BASED ON A THOROUGH REVIEW OF THE REPORTED COMPLAINT, BOSTON SCIENTIFIC HAS ASSIGNED AN INVESTIGATION CONCLUSION CODE OF CAUSE NOT ESTABLISHED.
IT WAS REPORTED THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE SHOCK THERAPY FOR ATRIAL FIBRILLATION (AF). THERAPY WAS EXHAUSTED. IT WAS NOTED THAT TWO SHOCKS TRIGGERED A CODE 1005, INDICATING AN OPEN CIRCUIT CONDITION. TECHNICAL SERVICES (TS) REVIEWED THE REPORTS AND PROVIDED TROUBLESHOOTING ACTIONS. THE DEVICE REMAINS IN USE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
IT WAS REPORTED THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE SHOCK THERAPY FOR ATRIAL FIBRILLATION (AF). THERAPY WAS EXHAUSTED. IT WAS NOTED THAT TWO SHOCKS TRIGGERED A CODE 1005, INDICATING AN OPEN CIRCUIT CONDITION. TECHNICAL SERVICES (TS) REVIEWED THE REPORTS AND PROVIDED TROUBLESHOOTING ACTIONS. THE DEVICE REMAINS IN USE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341990 | MOMENTUM EL ICD DR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) | LWS | BOSTON SCIENTIFIC CORPORATION | D121 | 240942 | 00802526587733 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 90 YR | Unknown | Other |