FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2192044 · Received August 4, 2011

Report

Report Number
6000001-2011-16601
Event Type
Malfunction
Date Received
August 4, 2011
Date of Event
July 14, 2011
Report Date
July 14, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP THAT EXPERIENCED A DAMAGED BATTERY ALARM WAS CONFIRMED BY BAXTER PERSONNEL DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM WAS DETERMINED TO BE DEPLETED BATTERIES RESULTING FROM USER ERROR. THIS CONDITION WAS RESOLVED ONSITE AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN BY REPLACING THE MAIN BATTERIES AND BATTERY HARNESS. THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.63.92. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ONSITE AT THE FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT A COLLEAGUE INFUSION PUMP EXPERIENCED A BATTERY ISSUE. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE EVENT IN ASSOCIATION WITH THIS CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.