FDA Adverse Event
Malfunction
Summary report: N
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
MDR report key: 2192033
·
Received August 4, 2011
Report
- Report Number
- 1030489-2011-01002
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Date of Event
- July 6, 2011
- Report Date
- July 6, 2011
- Manufacturer
- MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE SHAFT IS CRACKED AT THE CENTER OF THE JAW PIVOT PIN, WITH THE FIXED LOWER JAW BENT DOWNWARD. THE LOCATION, DIRECTION, AND AMOUNT OF FORCE REQUIRED IN ORDER INDUCE FRACTURE OF THE SHAFT IS CONSISTENT WITH APPLICATION OF EXCESSIVE FORCE, RESULTING IN THE FOREGOING EVENT. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THIS DEVICE DID NOT REVEAL ANY NON-CONFORMANCES TO SPECIFICATION OR DEVIATIONS IN PROCEDURE WHICH MIGHT CONTRIBUTE TO THE REPORTED EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SPINAL PROCEDURE AND THE TIP OF THE INSTRUMENT BROKE. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING | GZ10G063 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |