FDA Adverse Event Malfunction Summary report: N

ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

MDR report key: 2192033 · Received August 4, 2011

Report

Report Number
1030489-2011-01002
Event Type
Malfunction
Date Received
August 4, 2011
Date of Event
July 6, 2011
Report Date
July 6, 2011
Manufacturer
MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE SHAFT IS CRACKED AT THE CENTER OF THE JAW PIVOT PIN, WITH THE FIXED LOWER JAW BENT DOWNWARD. THE LOCATION, DIRECTION, AND AMOUNT OF FORCE REQUIRED IN ORDER INDUCE FRACTURE OF THE SHAFT IS CONSISTENT WITH APPLICATION OF EXCESSIVE FORCE, RESULTING IN THE FOREGOING EVENT. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THIS DEVICE DID NOT REVEAL ANY NON-CONFORMANCES TO SPECIFICATION OR DEVIATIONS IN PROCEDURE WHICH MIGHT CONTRIBUTE TO THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A SPINAL PROCEDURE AND THE TIP OF THE INSTRUMENT BROKE. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHOPEDIC MANUAL SURGICAL INSTRUMENT ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING GZ10G063

Patients

Seq Age Sex Outcome Treatment
1