FDA Adverse Event Malfunction Summary report: N

LEGEND FOOTED ATTACHMENT

MDR report key: 2192013 · Received August 4, 2011

Report

Report Number
1625507-2011-00102
Event Type
Malfunction
Date Received
August 4, 2011
Report Date
July 6, 2011
Manufacturer
MDT POWERED SURGICAL SOLUTIONS
Product Code
HBB
PMA / PMN Number
K020069
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORT CONFIRMED. EVALUATION DETERMINED THAT THE FOOTED PORTION WAS DAMAGE BY TOOL CONTACT. NO ADDITIONAL INFORMATION WAS AVAILABLE ON FOLLOW-UP. THE USER MANUAL CONTAINS THE FOLLOWING WARNING "THE ATTACHMENT SHOULD NOT BE USED IF ANY PART OF THE ATTACHMENT APPEARS TO BE BENT, LOOSE, MISSING, OR DAMAGED. EXCESSIVE PRESSURE OR IMPROPER HANDLING, SUCH AS BENDING OR PRYING, OF THE ATTACHMENT OR DISSECTING TOOL MAY CAUSE INJURY TO THE PATIENT, OPERATOR AND/OR OPERATING ROOM STAFF." OUR RECOMMENDATION FOR FACTORY SERVICE IS BASED ON THE FACILITY'S USAGE; HOWEVER IT SHOULD NOT EXCEED 24 MONTHS. THIS DEVICE HAS BEEN IN USE FOR 36 MONTHS WITHOUT ANY RECORD OF FACTORY SERVICE.

Description of Event or Problem · 1

DEVICE RETURNED FOR NON-SPECIFIED REPAIR. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO COMPLAINT ON DECONTAMINATION DUE TO THE ATTACHMENT BEING DAMAGED BY TOOL CONTACT. NO ADDITIONAL INFORMATION WAS AVAILABLE ON FOLLOW-UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LEGEND FOOTED ATTACHMENT MOTOR, DRILL, PNEUMATIC HBB MDT POWERED SURGICAL SOLUTIONS N/A N/A

Patients

Seq Age Sex Outcome Treatment
1