FDA Adverse Event
Summary report: N
TOTALCARE
MDR report key: 2191993
·
Received July 29, 2011
Report
- Report Number
- 1824206-2011-04014
- Date Received
- July 29, 2011
- Date of Event
- July 12, 2011
- Report Date
- July 12, 2011
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TECH FOUND THAT THE HEAD SECTION WOULD NOT RAISE. MANUAL WOULD NOT WORK EITHER. PROBLEM DUE TO BAD CYLINDER. REPLACED HEAD CYLINDER TO RESOLVE THIS ISSUE.
Description of Event or Problem · 1
COMPLAINT INFO INDICATED THAT THE HEAD SECTION WILL NOT RAISE. NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | P1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |