FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2191985 · Received July 22, 2011

Report

Report Number
1720753-2011-20322
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
July 11, 2011
Report Date
July 22, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE SVC REP RETRIEVED THE ERROR LOG FILES, AND RESEATED THE PRINTED CIRCUIT BOARDS AND THE CONNECTORS ON THE WORKSTATION. THE RTOS AND THE POWER SUPPLY WERE REPLACED. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS WOULD INTERMITTENTLY SHUTDOWN AND REBOOT. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1