ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05616
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Report Date
- May 6, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DISREGARD THE INITIAL 3500A REPORT (#2531779-2011-05616) SUBMITTED FOR THIS COMPLAINT. THE DESCRIPTION PROVIDED IN REPORT #2531779-2011-05616 WAS CORRECTLY SUBMITTED ON THE INITIAL 3500A REPORT #2531779-2011-05461.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6). (B)(4).
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. DURING EVALUATION THE FORCE SENSOR PINS WERE FOUND TO BE PARTIALLY DISLODGED. THIS REPORT IS BEING MADE BASED ON INVESTIGATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |