FDA Adverse Event
Malfunction
Summary report: N
CS100
MDR report key: 2191858
·
Received July 29, 2011
Report
- Report Number
- 2249723-2011-00114
- Event Type
- Malfunction
- Date Received
- July 29, 2011
- Date of Event
- May 23, 2011
- Report Date
- May 23, 2011
- Manufacturer
- DATASCOPE CORP.
- Product Code
- DSP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REP REPLACED THE DC TO AC INVERTER PRINTED CIRCUIT BOARD. THE UNIT WAS TESTED TO FACTORY SPEC. IT FUNCTIONED NORMALLY AND WAS RETURNED TO THE CUSTOMER. A REVIEW OF THE DEVICE'S SERVICE HISTORY DOES NOT INDICATE ANY SYSTEMIC ISSUES.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHILE THE UNIT WAS IN USE ON A PT, THE DISPLAY WENT BLANK. THE UNIT WAS REPLACED AND THERAPY WAS CONTINUED. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CS100 | INTRA-AORTIC BALLOON PUMP | DSP | DATASCOPE CORP. | CS100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |