FDA Adverse Event Malfunction Summary report: N

LEAD MODEL 302

MDR report key: 2191851 · Received July 22, 2011

Report

Report Number
1644487-2011-01667
Event Type
Malfunction
Date Received
July 22, 2011
Date of Event
June 26, 2011
Report Date
June 27, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS REPORTED BY A COMPANY REP THAT A VNS PT WAS SEEN AT AN OFFICE VISIT IN WHICH HIGH LEAD IMPEDANCE (DC DC 7) WAS IDENTIFIED. THE LAST KNOWN SETTINGS WERE FROM (B)(6) 2011. THE PT WAS REFERRED FOR X-RAYS AND THE GENERATOR WAS PROGRAMMED OFF. X-RAYS WERE RECEIVED AND REVIEWED BY THE MFR. REVIEW OF X-RAYS INDICATED THAT THE GENERATOR PLACEMENT APPEARED TO BE NORMAL AND THE FILTER FEED-THROUGH WIRES APPEARED TO BE INTACT. THE LEAD CONNECTOR PIN APPEARED TO BE FULLY INSERTED INTO THE GENERATOR CONNECTOR BLOCK. THE ELECTRODES APPEARED TO BE PLACED IN NORMAL ARRANGEMENT. MOST OF THE LEAD WAS FOUND TO BE COILED NEAR THE CHEST AREA IN A CONDITION EXHIBITED BY A "TWIDDLER." A LEAD DISCONTINUITY WAS FOUND IN THE CHEST AREA BASED ON THE COILING OF THE LEAD. AT THE MOMENT, GOOD FAITH ATTEMPTS TO OBTAIN ADD'L INFO HAVE BEEN UNSUCCESSFUL TO DATE.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE VNS PATIENT UNDERWENT GENERATOR AND LEAD REPLACEMENT SURGERY ON (B)(6) 2014 DUE TO LEAD DISCONTINUITY. THE EXPLANTING FACILITY DISCARDED THE EXPLANTED DEVICES; THEREFORE, NO ANALYSIS CAN BE PERFORMED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED FROM THE PHYSICIAN THAT NO PATIENT MANIPULATION OR TRAUMA OCCURRED THAT COULD HAVE CONTRIBUTED TO THE REPORTED HIGH LEAD IMPEDANCE. THE SURGERY FOR LEAD REVISION HAS NOT BEEN SCHEDULED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LEAD MODEL 302 LYJ CYBERONICS, INC. 302-20 009894

Patients

Seq Age Sex Outcome Treatment
1 31 YR