FDA Adverse Event Malfunction Summary report: N

ENTERALITE INFINITY ENTERAL FEEDING PUMP

MDR report key: 2191847 · Received July 29, 2011

Report

Report Number
1722139-2011-00107
Event Type
Malfunction
Date Received
July 29, 2011
Date of Event
June 19, 2011
Report Date
June 29, 2011
Manufacturer
MOOG MED DEVICES GROUP
Product Code
LZH
PMA / PMN Number
K031199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM TESTING FOR THIS PUMP WAS REVIEWED AND FOUND TO BE FUNCTIONING NORMALLY. ALL ALARMS FUNCTIONED ACCORDING TO SPECS. VOLUMETRIC ACCURACY WAS FOUND TO BE WITHIN SPEC. SEVERAL TESTS WERE RUN WITH PEDIASURE AT THE SETTINGS USED BY THE CUSTOMER. EACH TIME WHEN THE BAG WAS EMPTY, THE PUMP STOPPED AND ALARMED NO FOOD. THE CUSTOMER COMPLAINT COULD NOT BE DUPLICATED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE PUMP DOES NOT STOP WHEN THE BAG IS EMPTY. PT'S MOTHER STATED THEY USED PEDIASURE AT 40ML/HR FOR 720ML. MOTHER ALLEGED THE PUMP KEPT RUNNING AFTER THE FORMULA WAS GONE AND NO ALARM SOUNDED. SHE IMMEDIATELY TURNED OFF THE PUMP AND USED A BACK UP PUMP BEFORE AIR COULD BE PUMPED INTO THE PT. THE PT'S CONDITION OF 'FAILURE TO THRIVE' WAS NOT CHANGED BY THE EVENT AND NO INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTERALITE INFINITY ENTERAL FEEDING PUMP LZH MOOG MED DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1 3 YR