FDA Adverse Event Other Summary report: N

PHOTOTHERAPY CHAMBER

MDR report key: 21918 · Received May 1, 1995

Report

Report Number
21918
Event Type
Other
Date Received
May 1, 1995
Date of Event
April 24, 1995
Report Date
May 1, 1995
Manufacturer
ULTRALITE ENTERPRISES, INC.
Product Code
KGL
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
SD, US
Reporter Occupation
PHYSICAL THERAPIST

Narratives

Description of Event or Problem · 1

ULTRAVIOLET THERAPY CHAMBER WAS CLEANED BY A MAINTENANCE TECHNICIAN. ULTRAVIOLET A & B WAVE BULBS WERE REMOVED FOR CLEANING. WHEN BULBS WERE REPLACED THEY WERE NOT PUT IN THE CORRECT SLOTS THUS PTS WERE SUBJECTED TO INCORRECT PHOTOTHERAPY DOSAGES. FOUR PTS WERE GIVEN THERAPY BEFORE THE BULB MIXUP WAS DISCOVERED. THEY ALL RECEIVED 1ST DEGREE BURNS. THERE IS NO WARNING ON THE BULBS OR CHAMBER ADDRESSING THE POSSIBILITY OF BULB MIXUP. THERE ARE ONLY SMALL RED DOTS ABOVE THE RED SLOTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHOTOTHERAPY CHAMBER ULTRA VIOLET CHAMBER KGL ULTRALITE ENTERPRISES, INC. V4472-IV NA

Patients

Seq Age Sex Outcome Treatment
1 37 YR Other