FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2191758 · Received July 28, 2011

Report

Report Number
1824206-2011-03996
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 12, 2011
Report Date
July 12, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND IN "DOWN" STORAGE AREA. FOUND HEAD HI-LO DRIFTS DOWN SLOWLY DUE TO FAULTY EMERGENCY TREND VALVE. REPLACED EMERGENCY TREND VALVE TO RESOLVE THIS PROBLEM.

Description of Event or Problem · 1

COMPLAINT ALLEGED THAT THE HEAD HI-LO WOULD DRIFT DOWN SLOWLY. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE TOTAL CARE FNL HILL-ROM, INC. P1900

Patients

Seq Age Sex Outcome Treatment
1