FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2191749 · Received July 28, 2011

Report

Report Number
1824206-2011-03992
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
May 26, 2011
Report Date
May 26, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE SEAL FOR THE MANUAL CPR VALVE WAS DETERIORATED, DEFORMED AND STUCK IN THE PILOT VALVE INSIDE THE MANIFOLD. THE TECHNICIAN REMOVED THE REMAINING PIECES OF THE SEAL AND REPLACED THE CPR VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE HEAD SECTION WILL NOT GO DOWN USING MANUAL CPR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1