Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN ISOLATED THE ISSUE TO THE POWER CORD. HE REPLACED THE POWER CORD TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE TECHNICIAN FOUND THE GROUNDING WAS OPEN DURING A SAFETY ANALYSIS.