ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05608
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Report Date
- July 9, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED AN OUT OF CALIBRATION FORCE SENSOR, CORROSION ON THE FORCE SENSOR, AND A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS NOT DUPLICATED DURING TESTING. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED AND THE BATTERY CAP THREADS WERE FOUND TO BE STRIPPED. THE USER GUIDE INSTRUCTS THE PATIENT TO REPLACE THE BATTERY CAP AT LEAST ONCE PER YEAR AND THAT CRACKS, CHIPS, OR DAMAGE TO THE PUMP MAY IMPACT THE BATTERY CONTACT AND / OR THE WATERPROOF FEATURE OF THE PUMP.
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. (B)(6). RECALL 2531779-03/24/2010-003-R.
THE PATIENT REPORTED THAT THE PUMP WOULD NOT DETECT A CARTRIDGE; THE PATIENT REPORTEDLY TRIED THREE DIFFERENT CARTRIDGES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |