FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2191629 · Received July 20, 2011

Report

Report Number
1720753-2011-20241
Event Type
Malfunction
Date Received
July 20, 2011
Date of Event
July 8, 2011
Report Date
July 20, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE KILOVOLTS WERE CHECKED AND SET AND THE SYSTEM WAS CALIBRATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM'S KILOVOLTS WERE TOO HIGH. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1