FDA Adverse Event Malfunction Summary report: N

7700

MDR report key: 2191526 · Received July 27, 2011

Report

Report Number
9680959-2011-01615
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
July 13, 2011
Report Date
July 27, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. PARTS FOR THE TUBE WERE REPLACED. THE SYSTEM WAS WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM STARTED UP, BUT WOULD NOT EMIT FLUOROSCOPIC X-RAYS AND THE TEMPERATURE LIGHT ILLUMINATED. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7700 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7700

Patients

Seq Age Sex Outcome Treatment
1