FDA Adverse Event Malfunction Summary report: N

COAGUCHEK® XS SYSTEM

MDR report key: 2191326 · Received August 4, 2011

Report

Report Number
1823260-2011-04193
Event Type
Malfunction
Date Received
August 4, 2011
Date of Event
July 11, 2011
Report Date
August 29, 2011
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE EVENT OCCURRED IN (B)(6).

Description of Event or Problem · 1

CALLER STATES PATIENT TESTED 4.9 INR ON THE COAGUCHEK XS PLUS SYSTEM WHILE A COMPARISON LAB RETURNED AS 3.7 INR. PATIENT'S WARFARIN DOSE WAS HELD FOR ONE DAY. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT SYSTEM AND REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COAGUCHEK® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 20408212

Patients

Seq Age Sex Outcome Treatment
1 062 YR COLCHICINE| WARFARIN| FLUCLOXACILLIN