FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2191298
·
Received July 28, 2011
Report
- Report Number
- 2032227-2011-01849
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- July 14, 2011
- Report Date
- July 14, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
THE CUSTOMER STATED THAT SHE EXPERIENCED HIGH BLOOD GLUCOSE LEVELS FOR THE PAST 3 DAYS. THE CUSTOMER'S MOST RECENT BLOOD GLUCOSE WAS OVER 300 MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT FAILED THE HIGH PRESSURE TEST. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC MINIMED | MMT-522LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |