FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2191296 · Received July 28, 2011

Report

Report Number
2032227-2011-01846
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 13, 2011
Report Date
July 13, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND DUE TO A CORRODED MOTOR HOME SWITCH.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE REPLACEMENT TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-522LNAL

Patients

Seq Age Sex Outcome Treatment
1 65 YR