FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2191283 · Received July 28, 2011

Report

Report Number
2032227-2011-01847
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
July 13, 2011
Report Date
July 13, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE LEVELS FOR THE PAST TWELVE HOURS. THE REPORTED BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 424 MG/DL. THE CUSTOMER DIDN'T BELIEVE THE INSULIN PUMP WAS DELIVERING INSULIN ACCURATELY. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PUMP PASSED THE PRIME TEST, BUT FAILED THE HIGH PRESSURE TEST. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 73 YR UNOMEDICAL QUICK-SET PARADIGM INSULIN| INFUSION SET: MMT397, LOT 2209767.