LUXSTAR UNIT-MOUNT LIGHT
Report
- Report Number
- 3004517290-2011-00008
- Event Type
- Malfunction
- Date Received
- July 19, 2011
- Date of Event
- June 21, 2011
- Report Date
- June 22, 2011
- Manufacturer
- DENTAL EQUIPMENT LLC (DBA NEWBERG MFG)
- Product Code
- EAZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAS A THREE PRONG LOCKING MECHANISM WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB DURING ROUTINE OPERATOR MAINTENANCE. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE. DENTAL EQUIPMENT LLC INITIATED A RECALL ON (B)(4) 2011 TO ADD A TETHER TO THE LENS HEAT SHIELD/COVER TO KEEP IT FROM FALLING OFF THE DENTAL LIGHT IF IT IS NOT RE-INSTALLED PROPERLY.
A DOCTOR WAS PERFORMING ROUTINE MEDICAL TREATMENT TO A PT WHEN THE LENS HEAT SHIELD/HOLDER FROM HIS DENTAL LIGHT CAME OFF AND FELL DOWN TOWARDS THE FLOOR STRIKING THE PT IN THE FACE. THE LENS HEAT SHIELD HOLDER CAUSED A SMALL BURN TO THE PT'S FACE. NO MEDICAL TREATMENT WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LUXSTAR UNIT-MOUNT LIGHT | DENTAL OPERATING LIGHT | EAZ | DENTAL EQUIPMENT LLC (DBA NEWBERG MFG) | UL1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |