FDA Adverse Event Malfunction Summary report: N

LUXSTAR UNIT-MOUNT LIGHT

MDR report key: 2191282 · Received July 19, 2011

Report

Report Number
3004517290-2011-00008
Event Type
Malfunction
Date Received
July 19, 2011
Date of Event
June 21, 2011
Report Date
June 22, 2011
Manufacturer
DENTAL EQUIPMENT LLC (DBA NEWBERG MFG)
Product Code
EAZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAS A THREE PRONG LOCKING MECHANISM WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB DURING ROUTINE OPERATOR MAINTENANCE. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE. DENTAL EQUIPMENT LLC INITIATED A RECALL ON (B)(4) 2011 TO ADD A TETHER TO THE LENS HEAT SHIELD/COVER TO KEEP IT FROM FALLING OFF THE DENTAL LIGHT IF IT IS NOT RE-INSTALLED PROPERLY.

Description of Event or Problem · 1

A DOCTOR WAS PERFORMING ROUTINE MEDICAL TREATMENT TO A PT WHEN THE LENS HEAT SHIELD/HOLDER FROM HIS DENTAL LIGHT CAME OFF AND FELL DOWN TOWARDS THE FLOOR STRIKING THE PT IN THE FACE. THE LENS HEAT SHIELD HOLDER CAUSED A SMALL BURN TO THE PT'S FACE. NO MEDICAL TREATMENT WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LUXSTAR UNIT-MOUNT LIGHT DENTAL OPERATING LIGHT EAZ DENTAL EQUIPMENT LLC (DBA NEWBERG MFG) UL1000

Patients

Seq Age Sex Outcome Treatment
1