FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2191276
·
Received July 28, 2011
Report
- Report Number
- 3004209178-2011-82334
- Event Type
- Malfunction
- Date Received
- July 28, 2011
- Date of Event
- July 15, 2011
- Report Date
- July 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE CUSTOMER STATED HE RECEIVED A MOTOR ERROR WHEN HE CALLED PREVIOUSLY. THE CUSTOMER STATED THAT IT APPEARED THE INSULIN PUMP WAS REWINDING PROPERLY. ADVISED CUSTOMER TO CHECK HIS BLOOD GLUCOSE AND FOUND THAT HIS GLUCOSE LEVEL WAS LOW. INSTRUCTED CUSTOMER TO RUN A DISPLACEMENT TEST AND THE INSULIN PUMP FAILED THE TEST. INSTRUCTED TO CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | PARADIGM 6MM CATHETER| QUICK-SET 23| UNOMEDICAL INFUSION SET: MMT-399 |