ANEURX ABDOMINAL STENT GRAFT
Report
- Report Number
- 2953200-2011-01355
- Event Type
- Injury
- Date Received
- July 29, 2011
- Date of Event
- June 29, 2011
- Report Date
- June 29, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVAL, RESULTS: (ENDOLEAK). RESULTS AND CONCLUSIONS: (AORTIC NECK ANGULATION OF 50 DEGREES).
AN ANEURX ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN UNK SIZE ABDOMINAL AORTIC ANEURYSM, APPROX TEN AND ONE HALF YEARS AGO. CURRENT VESSEL MORPHOLOGY WAS REPORTED AS THE ANEURYSM IS 11CM; MILD TORTUOSITY AND CALCIFICATION IN THE VESSELS; A SHORTENED NECK; AN AORTIC NECK ANGULATION OF 50 DEGREES; AN AORTIC NECK DIAMETER AT THE RENAL ARTERIES OF APPROX 23 MM. IT WAS REPORTED THAT IN THE PAST, THE PT MUST HAVE HAD A DISTAL TYPE I ENDOLEAK IN THE RIGHT ILIAC THAT HAD EXPANDED WITH THE WHOLE ANEURYSM WHICH NECESSITATED AT LEAST ONE INTERVENTION BECAUSE, THE LIMB HAS BEEN RELINED WITH ANOTHER MFR'S EXTENSION CUFF. IT IS ASSUMED THAT THERE WAS PRIOR SAC EXPANSION, WITH NO ENDOLEAK, SO THEY RELINED THE LIMB DUE TO A TYPE V ENDOLEAK AND THEN WHEN THE ILIAC GOT BIGGER, THE OTHER MFR'S CUFF WAS PLACED. THE PROXIMAL SEAL IS MINIMAL, AT 6 MM BELOW THE RENALS. THE PHYSICIAN COIL EMBOLIZED THE RIGHT HYPOGASTRIC ARTERY AND PLACED AN ENDURANT LIMB FROM THE ANEURX TO EXTERNAL ILIAC ARTERY. THE PHYSICIAN DECIDED TO CORRECT THE IMMEDIATE PROBLEM ONLY, RATHER THAN CONVERTING TO A TALENT CONVERTER TO EXCLUDE THE ANEURX, IN ORDER TO STOP THE ANEURYSM FROM GROWING BECAUSE OF THE POROSITY OF THE OLD STENT GRAFT. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX ABDOMINAL STENT GRAFT | MIH | MEDTRONIC CARDIOVASCULAR | NA | M00J750068 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Required Intervention |