FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2191195 · Received July 25, 2011

Report

Report Number
1720753-2011-20382
Event Type
Malfunction
Date Received
July 25, 2011
Date of Event
July 13, 2011
Report Date
July 25, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE RESEATED THE CABLES AT THE C-ARM MODULE AND THE WORKSTATION. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD DISPLAY AN INTERMITTENT COMMUNICATION ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1