ANEURX ABDOMINAL STENT GRAFT
MDR report key: 2191181
·
Received July 29, 2011
Report
- Report Number
- 2953200-2011-01353
- Event Type
- Injury
- Date Received
- July 29, 2011
- Date of Event
- June 29, 2011
- Report Date
- June 29, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR SANTA IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). RESULTS: (ENDOLEAK). RESULTS AND CONCLUSIONS: (AORTIC NECK ANGULATION OF 70-80 DEGREES).
Description of Event or Problem · 1
AN ANEURX ABDOMINAL STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM, APPROX TWO YEARS AGO. VESSEL MORPHOLOGY WAS REPORTED AS MILD TO MODERATE CALCIFICATION AND TORTUOUSITY; AN AORTIC NECK ANGULATION OF 70-80 DEGREES; AN UNK AORTIC NECK DIAMETER. AT A RECENT FOLLOW-UP VISIT, THE CT SHOWED A PROXIMAL TYPE I ENDOLEAK, WITHOUT MIGRATION. THE PHYSICIAN PLACED AN ENDURANT CUFF AND ANOTHER MFR'S STENT, WHICH CORRECTED THE TYPE I ENDOLEAK. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX ABDOMINAL STENT GRAFT | MIH | MEDTRONIC CARDIOVASCULAR SANTA IRELAND | NA | V00323935 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention |