FDA Adverse Event Injury Summary report: N

ANEURX ABDOMINAL STENT GRAFT

MDR report key: 2191181 · Received July 29, 2011

Report

Report Number
2953200-2011-01353
Event Type
Injury
Date Received
July 29, 2011
Date of Event
June 29, 2011
Report Date
June 29, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA IRELAND
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: (ENDOLEAK). RESULTS AND CONCLUSIONS: (AORTIC NECK ANGULATION OF 70-80 DEGREES).

Description of Event or Problem · 1

AN ANEURX ABDOMINAL STENT GRAFT WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM, APPROX TWO YEARS AGO. VESSEL MORPHOLOGY WAS REPORTED AS MILD TO MODERATE CALCIFICATION AND TORTUOUSITY; AN AORTIC NECK ANGULATION OF 70-80 DEGREES; AN UNK AORTIC NECK DIAMETER. AT A RECENT FOLLOW-UP VISIT, THE CT SHOWED A PROXIMAL TYPE I ENDOLEAK, WITHOUT MIGRATION. THE PHYSICIAN PLACED AN ENDURANT CUFF AND ANOTHER MFR'S STENT, WHICH CORRECTED THE TYPE I ENDOLEAK. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX ABDOMINAL STENT GRAFT MIH MEDTRONIC CARDIOVASCULAR SANTA IRELAND NA V00323935

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention