HS III PROXIMAL SEAL SYSTEM 3.8MM
Report
- Report Number
- 2242352-2011-01028
- Event Type
- Injury
- Date Received
- July 29, 2011
- Date of Event
- July 7, 2011
- Report Date
- July 8, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
INSPECTION: VISUAL INSPECTION REVEALED THAT THE SEAL WAS RETURNED W/O THE LOADING DEVICE AND THE DELIVERY DEVICE. THE TENSION SPRING ASSEMBLY WAS REC'D SEPARATE WITH THE BLUE STRING CUT. THE SEAL WAS COMPLETELY UNRAVELED AND BLOODY. BASED UPON THE REC'D CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL DID NOT PROVIDE SUFFICIENT HEMOSTASIS" COULD NOT BE CONFIRMED. IT IS NOT POSSIBLE TO CONFIRM OR DENY THE COMPLAINT AS THE SEAL WAS REC'D COMPLETELY UNRAVELED SO NO FURTHER TESTING COULD BE DONE. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL DID NOT PROVIDE SUFFICIENT HEMOSTASIS. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY FURTHER PT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS III PROXIMAL SEAL SYSTEM 3.8MM | CLAMPLESS BEATING HEART | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3038 | 25018646 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |