FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2191163 · Received July 19, 2011

Report

Report Number
1720753-2011-20187
Event Type
Malfunction
Date Received
July 19, 2011
Date of Event
June 30, 2011
Report Date
July 19, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE DISPLAY ADAPTER PRINTED CIRCUIT BOARD WAS REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED SNOWY AND SUPERIMPOSED IMAGES ON THE MONITOR. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1