FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER ED 2

MDR report key: 2191144 · Received July 19, 2011

Report

Report Number
9615050-2011-00540
Event Type
Malfunction
Date Received
July 19, 2011
Date of Event
June 28, 2011
Report Date
June 28, 2011
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED UNRESTRICTED FLOW. THE DEVICE WAS RETURNED TO THE CENTRAL SUPPLY DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "CASSETTE TEST FAILURE." NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, UNRESTRICTED FLOW WAS NOTED WHEN THE CASSETTE WAS INSERTED INTO THE DEVICE AND THE DOOR WAS CLOSED. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE PUMP WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLUM A+ DRIVER ED 2 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE: LIST#20791, SN (B)(4)