COLONRING/COMPRESSION ANASTOMOSIS RING (CAR27)
Report
- Report Number
- 3005278776-2011-00036
- Event Type
- Injury
- Date Received
- July 29, 2011
- Date of Event
- June 7, 2010
- Report Date
- June 29, 2011
- Manufacturer
- NITI SURGICAL SOLUTIONS LTD.
- Product Code
- FZP
- PMA / PMN Number
- K062008
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
THIS REPORT IS SUBMITTED ON BEHALF OF THE MFR (NITI SURGICAL SOLUTIONS LTD; 3005278776) AND THE IMPORTER (B)(4), AS NITI SURGICAL SOLUTIONS LTD. SERVES AS THE DESIGNATED COMPLAINT HANDLING UNIT FOR BOTH FACILITIES. THE PRODUCTION HISTORY FILES WERE REVIEWED, INDICATING THAT THE DEVICE WAS RELEASED ACCORDING TO ITS SPECS. ANASTOMOTIC LEAKAGE, INCLUDING ABSCESS, IS ONE OF THE MOST COMMON COMPLICATIONS OF COLORECTAL ANASTOMOTIC PROCEDURES WITH NO RELATION TO THE METHOD USED FOR THE CREATION OF THE ANASTOMOSIS. RECTOSCOPIC EXAMS DURING THE FIRST MONTH FOLLOWING THE PROCEDURE, LOW ALBUMIN VALUES AND CHEMORADIATED TISSUE ARE CONSIDERED AS RISK FACTOR FOR POSTOPERATIVE LEAKAGE. THE CURRENT CUMULATIVE LEAK RATE ASSOCIATED WITH THE USE OF THE COLONRING DEVICE IS WITHIN THE RANGE REPORTED IN THE LITERATURE FOR OTHER METHODS. AN INDEPENDENT QUALITY AND SAFETY MONITORING COMMITTEE HAS FOUND THE RELATION OF THE EVENT TO THE DEVICE TO BE UNDETERMINED.
THE PT HAD AN OPEN LAR PROCEDURE FOR MALIGNANT RECTAL TUMOR. THE ANASTOMOSIS WAS CREATED WITH THE COLONRING DEVICE AND ILEOSTOMY WAS CONSTRUCTED. APPENDECTOMY PROCEDURE WAS ALSO PERFORMED WITHIN THE SAME OPERATION. RECTOSCOPIC EXAM BEFORE DISCHARGE REVEALED THAT THE RING IS SLIGHTLY DISLOCATED AND A SMALL INSUFFICIENCY. A DAY AFTER DISCHARGE THE PT RETURNED TO THE HOSPITAL WITH FEVER, WORSENING OF GENERAL CONDITION AND A HIGH OUTPUT STOMA. RECTOSCOPIC EXAM REVEALED THAT THE RING WAS COMPLETELY DISLOCATED AND LOT OF PUS. AFTER 5 RECTOSCOPIC EXAMS, THE DIAGNOSIS WAS ANASTOMOTIC INSUFFICIENCY AND AN ABSCESS. THE PT WAS TREATED WITH LAVAGE. THE RING WAS REMOVED ON (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLONRING/COMPRESSION ANASTOMOSIS RING (CAR27) | IMPLANTABLE CLIP (FZP) | FZP | NITI SURGICAL SOLUTIONS LTD. | COLONRING (CAR27) | 43001001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Required Intervention | METAMIZOL| PANTOZOL |