FDA Adverse Event Injury Summary report: N

VASOVIEW HEMOPRO EVH SYSTEM

MDR report key: 2191089 · Received July 29, 2011

Report

Report Number
2242352-2011-01034
Event Type
Injury
Date Received
July 29, 2011
Date of Event
June 29, 2011
Report Date
July 11, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WILL REPORTEDLY NOT BE RETURNED TO MAQUET CARDIAC SURGERY FOR INVESTIGATION. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCES WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VESSEL HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO INSULATION CAME OFF THE JAWS. THE TIP WAS RETRIEVED WITH NO ISSUES THROUGH THE ORIGINAL INCISION. THIS OCCURRED WHEN THE PROCEDURE WAS ALMOST COMPLETED SO NO REPLACEMENT WAS USED TO COMPLETE THE PROCEDURE. THE PRODUCT WAS DISCARDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASOVIEW HEMOPRO EVH SYSTEM ENDOSCOPIC VESSEL HARVESTING SYSTEM GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25034980

Patients

Seq Age Sex Outcome Treatment
1 NA Other