FDA Adverse Event
Malfunction
Summary report: N
CHEMETRON SERIES 4000
MDR report key: 2190989
·
Received August 1, 2011
Report
- Report Number
- 2190989
- Event Type
- Malfunction
- Date Received
- August 1, 2011
- Date of Event
- June 23, 2011
- Report Date
- August 1, 2011
- Manufacturer
- ALLIED HEALTHCARE PRODUCTS, INC.
- Product Code
- CAN
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MI, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
NITROGEN MANIFOLD NOT WORKING PROPERLY. MANIFOLD DELIVERS GAS VIA RESERVE SIDE, NOT PRIMARY SIDE. MAINLY SEEN ON LEFT BANK SIDE OF MANIFOLD. THIS MANIFOLD ALSO DISPLAYS A VISUAL AND AUDIBLE INDICATOR THAT GAS TANKS ARE EMPTY, HOWEVER THE MANIFOLD HAS NOT PNEUMATICALLY SWITCHED OVER YET. FALSE INDICATION TO USER WHO IDENTIFIES EMPTY CYLINDERS BY VISUAL AND AUDIBLE INDICATOR, AND EXTERNAL GAUGES, WHICH APPEAR TO BE AT 0.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CHEMETRON SERIES 4000 | REGULATOR, PRESSURE, GAS CYLINDER | CAN | ALLIED HEALTHCARE PRODUCTS, INC. | 4000 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |