HF SENSOR DELIVERY SYSTEM
Report
- Report Number
- 3004936110-2025-00791
- Event Type
- Injury
- Date Received
- April 25, 2025
- Date of Event
- April 22, 2025
- Report Date
- May 2, 2025
- Manufacturer
- ABBOTT MEDICAL
- Product Code
- MOM
- UDI-DI
- 05414734509237
- PMA / PMN Number
- P100045
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE INVESTIGATION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.
A REVIEW OF THE INFORMATION PROVIDED WAS PERFORMED, AND THE SENSOR WAS WITHIN THE TOLERANCE FOR CM MEAN; THEREFORE, THE REPORTED EVENT IS UNCONFIRMED. THIS REPORTED EVENT WAS INVESTIGATED FOR POSSIBLE INACCURATE READING. BASED ON A BACKEND LOG ANALYSIS, IT WAS CONFIRMED THAT THE SENSOR WAS OPERATING WITHIN THE EXPECTED FREQUENCY RANGE OF 30-37.5-MHZ. THE SENSOR WAS OPERATING AT 35.00 MHZ, 34.95 MHZ, AND 35.03 MHZ DURING THE REVIEW OF THE APPLICABLE READING(S). A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT. A LOT REVIEW WAS PERFORMED VIA COMPLAINT HANDLING SYSTEM (EPIQ) USING A SEARCH ON THE BATCH NUMBER. CONCLUSIVELY, THERE ARE NO ADDITIONAL COMPLAINT(S) RELATED TO THE ASSOCIATED BATCH, AND THE REPORTED ISSUE.
A RIGHT HEART CATHETERIZATION WAS PERFORMED TO CONFIRM THE VALIDITY OF THE PRESSURE SENSOR READINGS. THE SENSOR WAS RECALIBRATED AND THE MEAN WAS DECREASED BY 4.4 MMHG. READINGS WERE OBSERVED UNDER THE MANUFACTURER'S PATIENT CARE NETWORK DATABASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1276799 | HF SENSOR DELIVERY SYSTEM | SYSTEM, HEMODYNAMIC, IMPLANTABLE | MOM | ABBOTT MEDICAL | CM2000 | 10589764 | 05414734509237 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR | Unknown | Required Intervention |