REFLEX HYBRID REVISION DRIVER DRAW ROD SCREW EXTRA
Report
- Report Number
- 9617544-2011-00249
- Event Type
- Malfunction
- Date Received
- July 19, 2011
- Date of Event
- June 22, 2011
- Report Date
- June 22, 2011
- Manufacturer
- STRYKER SPINE BORDEAUX
- Product Code
- LXH
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
METHOD: VISUAL INSPECTION, COMPLAINT HISTORY ANALYSIS, MFG RECORD REVIEW, RISK REVIEW, PHOTOGRAPHS. RESULT: VISUAL INSPECTION CONFIRMS TIPS OF BOTH INSTRUMENTS ARE BROKEN. BROKEN FRAGMENTS NOT RETURNED. COMPLAINT HISTORY ANALYSIS - 48 COMPLAINTS HAVE BEEN RECEIVED RELATING TO DRAW ROD TIP DEFORMATIONS ON THIS DRIVER. INVESTIGATIONS INTO PAST COMPLAINTS CONCLUDED THAT THE FAILURES WERE THE RESULT OF USER ERROR I.E. OVER-ANGULATION OF DRAW ROD WHEN CONNECTED TO SCREW, INSUFFICIENT DEPTH OF SCREWING BETWEEN DRAW ROD AND SCREW. FOR FOUR OF THOSE COMPLAINTS, FAILURE WAS REPORTED TO HAVE OCCURRED DURING SCREW INSERTION WHICH SURGICAL TECHNIQUE EXPLICITLY WARNS NOT TO DO. MFG RECORD REVIEW - NO RELEVANT DEVIATIONS REPORTED. RISK REVIEW - DRAW ROD IS MADE OF BIOCOMPATIBLE STAINLESS STEEL TO MITIGATE RISK OF BROKEN TIPS REMAINING INSIDE THE PT. CONCLUSION: THE DEFECTS ARE BELIEVED TO BE THE RESULT OF USER-ERROR. WHILE THE REVISION DRIVER IS ATTACHED TO THE SCREW, PIVOTING OR ANGULATION OF THE INSTRUMENT SHOULD BE AVOIDED AS THIS CAN CAUSE THE DRAW ROD TIP TO BEND OR BREAK. IN ORDER TO AVOID THE RIST OF DRAW ROD TIP BREAKAGE, THE REVISION DRIVER MUST BE AXIALLY ALIGNED WITH THE SCREW TRAJECTORY AND FULLY SEATED IN THE SCREW HEAD BEFORE INSERTING OR TIGHTENING THE DRAW ROD. NO CAPA IS NECESSARY AT THIS TIME.
IT WAS REPORTED THAT "BROKE DURING SCREW REMOVAL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REFLEX HYBRID REVISION DRIVER DRAW ROD SCREW EXTRA | INSTRUMENT | LXH | STRYKER SPINE BORDEAUX | NA | 109526 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |