MINMED QUICK SET
Report
- Report Number
- 3003442380-2025-07187
- Event Type
- Malfunction
- Date Received
- April 25, 2025
- Date of Event
- March 26, 2025
- Report Date
- May 30, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018051
- PMA / PMN Number
- K160648
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES.
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6010413 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010413 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 PACKAGING IN THE MULTIVAC 12, ON 26/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 30/MAY/2025 AGAINST MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION, HARM CODE UNTREATED DIABETIC KETOACIDOSIS WHICH THE PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY AN HCP OR REQUIRES EMERGENCY ADVANCED LIFE SUPPORT TO PREVENT PERMANENT ORGAN DAMAGE (ELEVATED BLOOD GLUCOSE LEVEL AND SYMPTOMS E.G., MODERATE TO LARGE KETONES, NAUSEA, VOMITING, ABDOMINAL PAIN, CONFUSION) AND LOT 6010413 AND NO OTHER COMPLAINT HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT NUMBER, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION, HARM CODE AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION OR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN EVENT OF HIGH BLOOD GLUCOSE LEVEL OF 422 MG/DL ON (B)(6) 2025. THE PATIENT WAS HOSPITALIZED ON (B)(6) 2025 FOR LESS THAN TWENTY-FOUR HOURS. DURING HOSPITALIZATION, THE PATIENT TREATED WITH SYRINGE AND PUMP. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 476375 | MINMED QUICK SET | UNO QUICK-SET 110/9 SC1 MECA | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | MMT-396A | 6010413 | 05705244018051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |