COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-16501
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Date of Event
- July 6, 2011
- Report Date
- July 11, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS DEVICE WAS NOT RETURNED TO BAXTER FOR EVALUATION, HOWEVER, A BAXTER FIELD SERVICE TECHNICIAN REPAIRED THIS DEVICE AT THE CUSTOMER SITE. DEVICE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THE DEVICE WAS REPAIRED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. THE CONDITION WAS CONFIRMED THROUGH A REVIEW OF THE EVENT HISTORY. THIS COMPLAINT IS AN ANCILLARY OF (B)(4). THE CAUSE WAS NOT IDENTIFIED AS THE CONDITION WAS NOT RELATED TO THE CUSTOMER REPORTED CONDITION. NO FURTHER TESTING HAS BEEN COMPLETED AT THIS TIME AND NO REPAIRS HAVE BEEN PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SENT.
DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED TWO OCCURRENCES OF AN UPSTREAM OCCLUSION SET ALARM, WHICH INTERRUPTED DELIVERY. "THIS MAY HAVE BEEN A FALSE ALARM. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT." THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |