ANEURX AAADVANTAGE STENT GRAFT SYSTEM
Report
- Report Number
- 2953200-2011-01259
- Event Type
- Injury
- Date Received
- July 8, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 8, 2011
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, RESULTS: (ENDOLEAK, MIGRATION); (DISEASE PROGRESSION; ANGULATED NECK); (AORTIC NECK ANGULATED AT GREATER THAN 90 DEGREES). CONCLUSION: (DISEASE PROGRESSION; ANGULATED NECK); (AORTIC NECK ANGULATED AT GREATER THAN 90 DEGREES).
AN ANEURX BIFURCATED STENT GRAFT (REF MFR 2953200-2011-01259) WAS IMPLANTED FOR ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROX FIVE YEARS AGO. THE PT HAS BEEN VERY COMPLIANT WITH FOLLOW-UPS; A SONOGRAM APPROX 15 MONTHS AGO SHOWED NO PROBLEMS. A ROUTINE CT IN ONE MONTH AGO SHOWED THAT THE STENT GRAFT HAD MIGRATED DISTALLY WITH A PROXIMAL TYPE I ENDOLEAK DUE TO DISEASE PROGRESSION. THE TOP OF THE STENT GRAFT IS AT THE TOP OF THE AAA, ABOUT 2.5-3 CM BELOW THE RENAL ARTERIES. CURRENTLY, THE AAA IS 9.7 CM IN DIAMETER, AND THE AORTIC NECK IS 15-17 MM IN DIAMETER, 2.5-3 CM LONG, AND ANGULATED AT GREATER THAN 90 DEGREES. THE MIGRATION WAS ATTRIBUTED TO THE AORTIC NECK ANGULATION. AN INTERVENTION WAS PERFORMED 3 WEEKS AGO. THERE WAS ONLY ROOM TO PLACE ONE ENDURANT CUFF (REF MFR 2953200-2011-01260), WHICH WAS PLACED RIGHT AT THE RENALS. THE CUFF WAS BALLOONED MULTIPLE TIMES WITH A RELIANT BALLOON, BUT THERE WAS A VERY SMALL PROXIMAL TYPE I ENDOLEAK, THAT IS EXPECTED TO SELF-RESOLVE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, THE PT IS FINE, AND WILL BE MONITORED. AN INTERNAL REVIEW OF FILMS PRIOR TO THE TIME OF THE ENDURANT CUFF PLACEMENT SHOWS A HIGHLY ANGULATED NECK (>90DEG) WITH A PROXIMAL TYPE I ENDOLEAK AND A DISTAL MIGRATION OF THE ANEURX STENT GRAFT TO AT OR JUST BELOW THE AORTIC NECK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM | MIH | MEDTRONIC CARDIOVASCULAR | NA | A83741 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Required Intervention |