FDA Adverse Event Injury Summary report: N

BIPOL LEAD MODEL 300

MDR report key: 2190782 · Received July 26, 2011

Report

Report Number
1644487-2011-01678
Event Type
Injury
Date Received
July 26, 2011
Date of Event
May 25, 2010
Report Date
July 7, 2011
Manufacturer
CYBERONICS INC
Product Code
LYJ
PMA / PMN Number
P970003
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED BY A NURSE VIA AN X-RAY REPORT THAT A VNS PT HAD AN ASYMMETRY OF VOCAL CORDS, WITH THE LEFT CORD IN A PARAMEDIAN POSITION, SUGGESTING A POSSIBILITY OF VOCAL CORD PARALYSIS. THE NURSE INDICATED THAT DIAGNOSTICS WERE WITHIN NORMAL LIMITS. X-RAY IMAGES WERE TAKEN AND SENT TO MFR FOR REVIEW ON (B)(6) 2011. THE ELECTRODES APPEARED TO BE ALIGNED PROPERLY. NO GROSS FRACTURE OR LEAD DISCONTINUITIES COULD BE VISUALIZED ON THE LEAD PORTION ASSESSED. GENERATOR COULD NOT BE ASSESSED IN THE X-RAY DUE TO THE LACK OF IMAGES OF THE CHEST. BASED ON THE X-RAYS RECEIVED THERE WERE NO ANOMALIES OBSERVED. HOWEVER, DUE TO THE LACK OF IMAGES AVAILABLE, THE ENTIRE LEAD BODY AND GENERATOR COULD NOT BE ASSESSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BIPOL LEAD MODEL 300 LYJ CYBERONICS INC 300-20 15270C

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention