FDA Adverse Event Malfunction Summary report: N

HAMILTON-C1

MDR report key: 21907279 · Received April 25, 2025

Report

Report Number
3001421318-2025-00560
Event Type
Malfunction
Date Received
April 25, 2025
Date of Event
April 10, 2025
Report Date
February 18, 2026
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
07630002800747
PMA / PMN Number
K181216
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG REF NR: (B)(4).

Additional Manufacturer Narrative · 0

INVESTIGATION OUTCOME: COMPLAINT DESCRIPTION: WHEN HOSPITAL STAFF WERE USING THIS C1 ON A PATIENT, TECHNICAL EVENT: 232005 APPEARED ON THE SCREEN, AN ALARM SOUNDED, AND IT SWITCHED TO AMBIENT MODE, SO THEY REPLACED IT WITH ANOTHER VENTILATOR. NO HEALTH CONSEQUENCES OR IMPACT. THE HAMILTON C1 WAS CHECKED BY THE TECHNICIAN ON SITE. A STRANGE NOISE COULD BE HEARD COMING FROM THE FAN AND THE FAN WAS IDENTIFIED THE SOURCE OF THE ISSUE. THIS CASE IS CONSIDERED AS CLOSED.

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: "WHEN HOSPITAL STAFF WERE USING THIS C1 ON A PATIENT, TECHNICAL EVENT: 232005 APPEARED ON THE SCREEN, AN ALARM SOUNDED, AND IT SWITCHED TO AMBIENT MODE, SO THEY REPLACED IT WITH ANOTHER VENTILATOR." THE PATIENT WAS NOT HARMED, AND NO CONSEQUENCES OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1227428 HAMILTON-C1 HAMILTON-C1 CBK HAMILTON MEDICAL AG 161001 07630002800747

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown