HAMILTON-C1
Report
- Report Number
- 3001421318-2025-00560
- Event Type
- Malfunction
- Date Received
- April 25, 2025
- Date of Event
- April 10, 2025
- Report Date
- February 18, 2026
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 07630002800747
- PMA / PMN Number
- K181216
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
HAMILTON MEDICAL AG REF NR: (B)(4).
INVESTIGATION OUTCOME: COMPLAINT DESCRIPTION: WHEN HOSPITAL STAFF WERE USING THIS C1 ON A PATIENT, TECHNICAL EVENT: 232005 APPEARED ON THE SCREEN, AN ALARM SOUNDED, AND IT SWITCHED TO AMBIENT MODE, SO THEY REPLACED IT WITH ANOTHER VENTILATOR. NO HEALTH CONSEQUENCES OR IMPACT. THE HAMILTON C1 WAS CHECKED BY THE TECHNICIAN ON SITE. A STRANGE NOISE COULD BE HEARD COMING FROM THE FAN AND THE FAN WAS IDENTIFIED THE SOURCE OF THE ISSUE. THIS CASE IS CONSIDERED AS CLOSED.
THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: "WHEN HOSPITAL STAFF WERE USING THIS C1 ON A PATIENT, TECHNICAL EVENT: 232005 APPEARED ON THE SCREEN, AN ALARM SOUNDED, AND IT SWITCHED TO AMBIENT MODE, SO THEY REPLACED IT WITH ANOTHER VENTILATOR." THE PATIENT WAS NOT HARMED, AND NO CONSEQUENCES OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1227428 | HAMILTON-C1 | HAMILTON-C1 | CBK | HAMILTON MEDICAL AG | 161001 | 07630002800747 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |