FDA Adverse Event Malfunction Summary report: N

POSEY STAYSAFE BED

MDR report key: 2190683 · Received July 29, 2011

Report

Report Number
2020362-2011-00241
Event Type
Malfunction
Date Received
July 29, 2011
Report Date
May 24, 2011
Manufacturer
J. T. POSEY CO.
Product Code
FNL
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL RESULTS: EVAL OF THE RETURNED PRODUCT FOUND THAT THE PANEL SIDE A WINDOW ZIPPER SLIDER BODY IS OPEN. PANEL SIDE C LOWER LEFT LEG ELASTIC IS BROKEN. NOTE: POSEY INSTRUCTIONS FOR USE HAS A WARNING STATEMENT: NEVER TRY TO RIP A PANEL OPEN AS THIS MAY DAMAGE THE ACCESS PANEL OR THE ZIPPER SLIDER BY BENDING IT OPEN. NEVER USE THE BED IF A ZIPPER SLIDER IS BENT OPEN OR DAMAGED, AS THIS MAY PREVENT THE ZIPPER FROM CLOSING SECURELY. NEVER LEAVE THE PT'S BEDSIDE UNTIL ALL ACCESS PANEL ZIPPERS ARE SECURELY CLOSED. TEST THE ENTIRE LENGTH OF EACH ZIPPER BY PRESSING AGAINST THE PANEL NEAR THE ZIPPER TO MAKE SURE IT IS SECURELY CLOSED AND THE ACCESS PANEL WILL NOT OPEN WHEN PRESSURE IS APPLIED. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SIDE AND FOOT PANELS HAVE ZIPPER DAMAGE. THE CUSTOMER COULDN'T PROVIDE MORE INFO ON THE SPECIFIC TYPE OF DAMAGE THERE IS TO THE ZIPPERS. THERE WAS NO PT INCIDENT OR INJURY REPORTED. INSPECTION OF THE RETURNED PRODUCT FOUND THAT THE PANEL SIDE A ZIPPER SLIDER BODY IS OPEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POSEY STAYSAFE BED FNL J. T. POSEY CO. 8060 NA

Patients

Seq Age Sex Outcome Treatment
1 NI